Fully Auto 5-Part Hematology Analyzer
ใบรับแจ้งรายการละเอียด เลขที่ 67-2-2-1-0000408
*อ่านคำเตือนในฉลากและเอกสารกำกับเครื่องมือแพทย์ก่อนใช้*
Warnings
* This analyzer can only be operated by test professionals, doctors or laboratory technicians who have been trained by Sinseng or its distributors.
• It is important for the hospital or organization that employs this equipment to carry out a reasonable service/maintenance plan. Neglect of this may result in machine breakdown or injury of human health.
• Be sure to operate the analyzer under the situation specified in this manual; otherwise, the analyzer will not work normally and the analysis results will be unreliable, which would damage the analyzer components and cause personal injury.
• The installation, authorization, upgrade and modification of the analyzer software must be performed by Sinseng-authorized personnel.
• There should be enough room on and below the countertop to accommodate the reagents and waste containers.
• The diluent container shall be put within 1.0m under the analyzer, lyse containers are placed inside the analyzer.
• The countertop (or the floor) where the analyzer is placed shall be able to withstand at least 40kg of weight.
• Make sure the analyzer is properly grounded.
• Before turning on the analyzer, make sure the input voltage meets the requirements.
• Do not contact the patients’ sample blood directly.
• Be sure to dispose of reagents, waste, samples, consumables, etc. according to government regulations.
• The reagents are irritating to eyes, skin and mucosa. Wear proper personal protective equipment (e.g., gloves, lab coat, etc.) and follow safe laboratory procedures when handling them and the contacted areas in the laboratory.
• If reagents accidentally spill on your skin or into your eyes, rinse the area with plenty of clean water and seek medical attention immediately.
• The sample probe tip is sharp and may contain biohazardous materials. Exercise caution to avoid contact with the probe when working around it.
• The sample may spill from the uncapped collection tubes and cause biohazard. Exercise caution to the uncapped collection tubes.
Precautions
• Probe Obstruction Carefully check reagents and samples and make sure they do not contain insoluble floating substance such as cellulose and protein fibrin in case the probes may be blocked.
• Water Quality should meet Class 2 national standards for laboratory water, otherwise damage to valve and pump as well as difficulty in cleaning can be resulted.
• Device Connection (1) For a device not permanently connected, please do not place it at a location that is hard to disconnect. (2) For all the external switches or breakers and external over-current protection device,
it is recommended to place them near the analyzer. (3) Devices connected with the network port of the analyzer should conform to the requirements of National Standards GB4793 of China
Contraindications :N/A